Semi-automatic biochemical analyzer BSA 3000

Convenient and reliable semi-automatic biochemical analyzer BSA 3000 with open system for reagents

Features:

Open system for reagents

Wavelength range  from 340 nm to 630 nm

Measurement accuracy: 0.0001 ABS

The optical system is protected from vibration

Methods of calculation of measurements: end point, kinetics, fixed time, absorption

Dichromatic testing is available

Calibration type:  linear and nonlinear

Flowing ditch

Graphical representation of the course of kinetic reactions

Simple management:

Easy installation and clear parameters allow operators to quickly program programs. The laboratory assistant can monitor in real time the reaction curves and temperature changes. Memory capacity - up to 3000 results per day. At any time, the data is available for printing and transmission through the LIS.

User-friendly interface:

Large touch LCD screen, simple interface for comfortable and easy use. BSA 3000 comes with a built-in thermal printer for stand-alone operations. There is a visual and audible signal to notify the operator of any errors.

Flexible and practical:

This French device is an open system and is compatible with all reagents and tests. Supports bichromatic testing for methods: kinetics, endpoint, fixed time. The temperature control option allows a quick change between three temperatures (25 °, 30 ° and 37 ° C) for optimal analysis.

Parameters to be set:

Measurement technique / Reaction type / Wavelength / Absorption limit / Temperature / Aspiration volume / Reagents / Standards / Sample form / Delay time / Measurement time / Units / Linearity control

Benefits:

The wide photometric range of the BSA 3000, high resolution and very low drift ensure accurate and reliable results.

Prevention of contamination and cross-contamination by aspiration volume (400 μl recommended for 32 μl measured volume) and wash cycles. Comprehensive quality control program, including Levi-Jennings charts  provides high quality results.

Simple workflow:

The device stores up to 90 different test programs available to the operator at any time. 26 tests were programmed under SFRI reagents for direct testing. The operator can program up to 64 individual tests, as well as create lists for effective work. Linear and non-linear calibration of curves is maintained in order to avoid unnecessary re-calibration in case of high cost of certain reagents (for turbidimetry).

BSA-3000-f04_edited.jpg

Reliability and reproducibility:

Optics and automatic filter selection provide reliable and reproducible results. The calculation is performed automatically, so that the results are displayed directly in the desired units.

BSA-3000-f02.jpg

Specification of the BSA 3000 biochemical analyzer

Method of measurement

Endpoint, kinetics, fixed time, absorption, sample form, reagent form, monochromatic and bichromatic modes

Light

Quartz Halogen Lamp 12/20 W. Light scattering <1.0% at 340 nm.

Optics

8 filters:
7 standard filters: 340 nm, 405 nm, 492 nm, 510 nm, 546 nm,

578 nm,  630 nm and one additional. Bandwidth <8 nm

Photometric range

Measuring range: 0.0000 to 3.0000 ABS

Sensitivity

0.0001 ABS   Drift: 0.002 ABS

Flowing ditch

Stainless steel with quartz window, Optical path: 10 mm, Volume measurement: 32 μl, Aspiration volume: 200 - 2000 μl

Calibration

Linear, nonlinear up to 6 points, factor

The amount of memory

3000 results daily

Thermostatic control (optional)

With Peltier elements 25 °, 30 °, 37 ° C, accuracy: ± 0.1 ° C

Input / Output ports

RS232 port for unidirectional LIS

Printer

Built-in thermal printer,

57.5 mm wide paper, 48 mm wide write width

Display

6 "touch monochrome display

Operating conditions

Temperature 15 ° C - 30 ° C, humidity 20% - 80%

Power grid

110/220 V ± 10%; 50 - 60 Hz

Dimensions

445 (W) x 190 (H) x 420 (D) mm

Weight

9 kg