top of page
69998-13893571_edited.jpg

Research menu:

Measured : 

WBC, RBC, HGB, PLT, MCV, MPV

Calculated : 

Lym #, Mon #, Gran #, HCT, MCH, MCHC, PCT%

Given in the form of histograms : 

Lym%, Mon%, Gran%, RDW-SD, RDW-CV, RDW%, P-CLR%

Additional Information:

Input power: 400 W

Dimensions: 330 (W) x 430 (H) x 380 (D) mm

Weight: 17.5 kg

Requirements for power grids:

AC 110/220 V ± 10%; 50-60 Hz

Automatic ESR analyzer

ESR 3000, SFRI

Reliable, economical and simple device

Bandwidth: Simultaneous processing of 30 random access samples

30 min or 60 min according to the modified Westegren method

             SFRI vacuum glass test tube

   The total sample volume is 1.28 ml

   Contains a solution of sodium citrate 0.32 ml

   Test tubes complete with labels

Photometric infrared registration principle
Automatic measurement of erythrocyte levels every 3 minutes

Signal input / output: ports for keyboard, mouse and barcode reader, interface for external printer and monitor,  4 USB ports, 1 RS232 port, 1 VGA port, 1 Ethernet port

Printing: fast thermal printer, paper width 57.5 mm, recording field width 48 mm.  External printer is an option

Display:  600 x 800 color graphics screen (8 (), available in English, French, Spanish and Russian

Memory capacity:   20,000 records, one-way LIS

Operating conditions: air temperature 18 ° С - 35 ° С.,
relative humidity <85% (maximum relative humidity), atmospheric pressure 86.0 - 106 kPa

Analytical characteristics:

Parameters: 

Linearity range: 

Accuracy (CV%): 

WBC (109 / L)

0 - 99.99

≤ 2.5% @ 5 - 10

RBC (1012 / L)

0 –9.99

≤ 2.0% @ 4 - 6

HGB (g / L)

0 –300

≤ 2.0% @ 50 - 200

MCV (fL)

40 –150

≤ 1.0% @ 70 - 120

PLT (109 / L)

10 –999

≤ 5.0% @ 200 - 450