top of page
HEMIX-H18-Light-f01.jpg

Research menu:

Measured : 

WBC, RBC, HGB, PLT, MCV, MPV

Calculated : 

Lym #, Mon #, Gran #, HCT, MCH, MCHC, PCT%

Given in the form of histograms : 

Lym%, Mon%, Gran%, RDW-SD, RDW-CV, RDW%, P-CLR%

Additional Information:

Input power: 400 W

Dimensions: 330 (W) x 430 (H) x 380 (D) mm

Weight: 17.5 kg

Requirements for power grids:

AC 110/220 V ± 10%; 50-60 Hz

Hematology analyzer

H18 Light, 3-Diff

Compact and economical solution

for low power laboratories

Bandwidth:  60 seconds for example

Sample volume: whole blood 9.6 μl,

               pre-diluted blood 20 μl

Quality control: quality control database, Levi-Jennings graphs, average SD calculation, CV, up to 9 control files for 31 tests

Impedance method for cell count determination and photometric measurement of hemoglobin by cyanide-free method

Signal input / output: ports for keyboard, mouse and barcode reader, interface for external printer and monitor,  4 USB ports, 1 RS232 port, 1 VGA port, 1 Ethernet port

Printing: fast thermal printer, paper width 57.5 mm, recording field width 48 mm.  External printer is an option

Display:  600 x 800 color graphics screen (8 (), available in English, French, Spanish and Russian

Memory capacity:   20,000 records, one-way LIS

Operating conditions: air temperature 18 ° С - 35 ° С.,
relative humidity <85% (maximum relative humidity), atmospheric pressure 86.0 - 106 kPa

Analytical characteristics:

Parameters: 

Linearity range: 

Accuracy (CV%): 

WBC (109 / L)

0 - 99.99

≤ 2.5% @ 5 - 10

RBC (1012 / L)

0 –9.99

≤ 2.0% @ 4 - 6

HGB (g / L)

0 –300

≤ 2.0% @ 50 - 200

MCV (fL)

40 –150

≤ 1.0% @ 70 - 120

PLT (109 / L)

10 –999

≤ 5.0% @ 200 - 450

bottom of page